Performance Specification 9

Specifications and Test Procedures for Gas Chromatographic Continuous Emission Monitoring Systems in Stationary Sources

40 CFR Part 60, Appendix B

Read the method on the eCFR →

From the method — regulatory text

Sections 1.0 and 2.0 of Performance Specification 9—Specifications and Test Procedures for Gas Chromatographic Continuous Emission Monitoring Systems in Stationary Sources, as published at 40 CFR Part 60, Appendix B. Retrieved 10/07/2026. The full method is on the eCFR.

1.0 Scope and Application

1.1 Applicability. These requirements apply to continuous emission monitoring systems (CEMSs) that use gas chromatography (GC) to measure gaseous organic compound emissions. The requirements include procedures intended to evaluate the acceptability of the CEMS at the time of its installation and whenever specified in regulations or permits. Quality assurance procedures for calibrating, maintaining, and operating the CEMS properly at all times are also given in this procedure.

2.0 Summary of Performance Specification

2.1 Calibration precision, calibration error, and performance audit tests are conducted to determine conformance of the CEMS with these specifications. Daily calibration and maintenance requirements are also specified.

EssentialRegs notes

Our plain-English reading of Performance Specification 9. Not the method text.

What it measures

Whether a continuous emission monitoring system that uses a gas chromatograph is acceptable for measuring specific gaseous compounds, based on its installation, performance and accuracy.

How it works

The system is installed at a representative location and operated through a test period. Its calibration is checked with gas standards, and its relative accuracy is determined by comparing its results with simultaneous results from a reference test method. The system is acceptable when it meets the limits in the specification.

Equipment

The gas chromatograph-based monitoring system being certified, calibration gas standards, and the reference method sampling equipment used for the relative accuracy test.

When a rule cites it

Rules that require a continuous emission monitoring system (CEMS) to demonstrate compliance, report emissions or track an emission limit on an ongoing basis. The system must be certified under the specification before its data count for those purposes, and recertified when the rule or a major system change calls for it.

Notes for compliance staff

A gas chromatograph reports individual compounds, which suits rules that limit specific organic compounds rather than a total. Performance Specification 8 is the specification for volatile organic compound monitors that report a group total.